NEW 5 4/5 Digits Hand Held DIGITAL MULTIMETER
50000/500,000 counts
|
|
|
Features :
Record MAX,
MIN, MAX-MIN reading. Crest (Instantaneous Peak hold) MAX, MIN, MAX- MIN
reading. Relative zero mode. 500,000 counts high resolution stable reading
mode.
Backlighted display, dBm readings.T1-T2 differential temperature readings
(KM859CF only). %4-20mA loop current readings. High noise rejection
filtered Line Level Frequency mode. Data Hold. Audible & Visible input
warning.
|
DISPLAY : 4-4/5 digits 50,000 counts.
Selectable stable mode 5-4/5 digits 500,000 counts for DC Voltage, & 6
digits 999,999 counts for frequency.
POLARITY : Automatic
UPDATE RATE : 4-4/5digits fast mode : 5 per second nominal; 5-4/5
digits stable mode : 1.25 per second nominal; 42 Segments Bar Graph : 60
per second max
Low battery indication
Data hold Function
LOW BATTERY : Below approx. 7V
OPERATING TEMPERATURE : 0°C to 45°C, £ 80% R.H.
STORAGE TEMPERATURE : -20°C to 60°C, 80% R.H. (With battery
removed)
SENSING : AC, AC+DC True RMS
POWER SUPPLY : Single alkaline 9V battery.
POWER CONSUMPTION : 5 mA typical
APO TIMING : Idle for 4.5 minutes
APO CONSUMPTION : 20 m A
DIMENSION : (L) 186mm X (W) 87mm X (H) 35.5mm; (L) 198mm X (W) 97mm
X (H) 55mm with holster
WEIGHT : 390 gm; 500 gm with holster Approx.
SAFETY : The Meters meet the requirements for double insulation to
IEC61010-1 (1995), EN61010-1 (1995), UL3111-1(6.1994), CSA C22.2 NO.
1010-1-92 Installation CategoryIII 1000 Volts ac and dc,CAT IV 600V DC &
AC.
E.M.C. : Meets EN55022(1994/A1; 1995/Class B) and EN50082-1(1992).
|
|
OVERLOAD PROTECTION
ACCURACY:
m A & mA : 0.63A/500V, IR 100kA, F fuse for
KM857;1A/600V, IR 100kA, F fuse for KM859CF
A : 12.5A500V, IR 10kA, F fuse for KM857;15A/1kV, IR
10kA F fuse for KM859CF
V : 1050Vrms, 1450Vpeak; mV : W &
Others : 600VDC/AC rms for KM857 (1050Vrms, 1450Vpeak for KM859CF
TRANSIENT PROTECTION : 8kV (1.2/50 m s surge)
STANDARD ACCESSORIES : Test leads (pair), holster,
battery installed, user's manual, banana plug K type Thermocouple
x1(KM859CF only), Tilt stand.
OPTIONAL ACCESSORIES : PC interface Kit (RS232
optical adapter cable + RS232 Software floppies)
ELECTRICAL SPECIFICATION
ACCURACY:
± (%Reading + number of digits) or otherwise specified, at 23°C ± 5°C &
less than 75% relative humidity. True RMS voltage & current accuracies are
specified from 5 % to 100% of range or otherwise specified, Maximum Crest
Factor < 5:1 at full scale & <10:1 at half scale, and with frequency
component within the specified frequency bandwidth for non-sinusoidal
waveforms.
|
DC VOLTAGE |
|
Range |
Accuracy
|
| |
KM859CF
|
KM857
|
| |
Accuracy
|
|
500.00mV, 5.0000V,
50.000V |
0.02%rdg+2dgts |
0.03%rdg + 2dgts |
| 500.00V |
0.04%rdg+2dgts |
0.05%rdg+2dgts |
| 1000.0V |
0.05%rdg+2dgts |
0.1%rdg+2dgts |
|
NMRR :
> 60dB @ 50/60Hz
CMRR : > 120dB @ DC, 50/60Hz, Rs = 1k
Input Impedance : 10M W , 30pF nominal (80pF
nominal
for 500mV ranges) |
|
|
CAPACITANCE
|
| Range |
Accuracy* |
| 50.000 nF |
0.8%+ 3d |
| 50.000 nF |
0.8%+ 3d |
| 50.000 nF |
1.0%+ 3d |
| 50.000 nF |
2.0%+ 3d |
| 50.000 nF |
3.5%+ 5d |
| 9999 m F |
5.0% + 5d |
|
*Accuracies with film capacitor or better
|
|
|
|
|
|
|
DIODE TESTER |
|
Range |
Accuracy
|
Test
Current (Typical ) |
Open
Circuit Voltages |
|
5.0000V |
1%+ 1d
|
0.8mA
|
<3.5 VDC
|
|
|
dBm : At 600
,
-11.76dBm to 54.25dBm,
Accuracy : ± 0.25dB + 2d (@40Hz -- 20kHz)
Input Impedance : 10M W , 30pF nominal
Selectable reference impedance values of 4, 8, 16, 32, 50, 75,
93, 110, 125, 135, 150, 200, 250, 300, 500, 600, 800, 900, 1000,
1200
Crest mode (Instantaneous Peak Hold) :
Accuracy : Specified accuracy ± 100 digits for changes
> 0.8ms in duration
|
|
|
~ Hz Line Level
Frequency |
| Range |
Accuracy* |
| 50.000 nF |
0.8%+ 3d |
| 50.000 nF |
0.8%+ 3d |
|
Sensitivities
(Sine RMS):
0.1V min @500mV range
T 1V min @5V range
10V min @50V range
100V min @500V range
900V min @1000V range |
|
|
|
AC & AC + DC CURRENT |
|
Range
|
KM859CF
|
KM857
|
Burden Voltage
|
| |
Accuracy
|
|
50Hz - 60Hz
|
0.5% + 50d
|
1.0% + 40d |
0.15 mV/ m A
|
|
500.00 m A
|
0.15 mV/ m A
|
|
5000.0 m A
|
3.3 mV/mA
|
|
500.00 mA
|
3.3 mV/mA
|
|
5.0000 A
|
0.03 V/A
|
|
10.000 A*
|
0.03 V/A
|
|
40Hz - 1kHz
|
|
|
|
|
500.00 m A
|
0.7% + 50d
|
1.0% + 40d
|
0.15 mV/ m A
|
|
5000.0 m A
|
0.15 mV/ m A
|
|
50.000 mA
|
3.3 mV/mA
|
|
500.00 mA
|
3.3 mV/mA
|
|
5.0000 A
|
0.03 V/A
|
|
10.000 A*
|
0.03 V/A
|
|
1kHz - 10kHz
|
|
|
|
| 500.00 m A |
|
|
0.15 mV/ m A |
| 5000.0 m A |
|
|
0.15 mV/ m A |
| 50.000 mA |
|
|
3.3 mV/mA |
| 500.00 mA |
2.0% + 50d
|
Unspec'd
|
3.3 mV/mA |
| 5.0000 A 10.000 A* |
Unspec'd
|
Unspec'd
|
0.03 V/A |
|
|
*10A continuous, 20A for 30 second max with 5 minutes cool down
interval
|
|
|
OHMS
|
|
Range |
KM859CF |
KM857 |
| |
Accuracy
|
|
500.00
 |
0.07% + 10d |
0.1% + 6d
|
5.0000 k |
0.07% +2d
|
50.000 k
 |
500.00 k
 |
5.0000 M
 |
0.2% + 6d |
0.4% + 6d |
50.000 M
 |
2.0% +6d |
2.0% + 6d |
| |
|
|
|
|
Open Circuit Voltage : < 1.3 VDC (<3VDC for 500 W ranges)
|
| Hz
LOGIC LEVEL FREQUENCY |
| Range
|
Accuracy
|
|
5.0000Hz- 2.00000MHz |
0.002%+4d
|
|
|
Sensitivities : 2.5Vp square wave
|
|
% DUTY CYCLE
|
| Range
|
Accuracy
|
|
0.1% -- 99.99% |
3d/kHz + 2d
|
|
|
Input Frequency : 5Hz - 500 kHz, 5V Logic family
|
|
|
|
T1-T2 DUAL
TEMPERATURE (KM859CF only) |
| Range
|
Accuracy
|
| -50.0 °C TO 1000.0°C |
0.3% + 1°C |
|
-58.0 °F TO 1832.0°F |
0.3% + 2°F
|
|
|
Thermocouple range & accuracy not included
|
|
% DUTY CYCLE |
| Range
|
Accuracy
|
|
0.1% -- 99.99% |
3d/kHz + 2d
|
|
|
Input Frequency : 5Hz - 500 kHz, 5V Logic family
|
|
DC CURRENT
|
| Range
|
Accuracy
|
Burden
Voltage |
| 500.00 m A |
0.15% rdg + 20dgts
|
0.15 mV/ m A |
| 5000.0 m A |
0.1% rdg + 20dgts |
0.15 mV/ m A |
| 50.000 mA |
0.15% rdg + 10dgts |
3.3 mV/mA |
| 500.00 mA |
0.1% rdg + 20dgts |
3.3 mV/mA |
| 5.0000 A |
0.5% rdg + 10dgts |
0.03 V/A |
| 10.000 A* |
0.5% rdg + 20dgts |
0.03 V/A |
|
|
*10A continuous, 20A for 30 second max with 5 minutes cool down
interval
DC Loop Current %4--20mA
4mA = 0% (zero); 20mA = 100% (span)
Resolution : 0.01% Accuracy : ± 25d
))) Audible Continuty Tester
Audible threshold : between 20 W and 120 W .
Response time
< 100 m s
|
|
|
AC & AC + DC VOLTAGE
|
|
Range
|
KM859CF
|
KM857
|
| |
Accuracy
|
|
20Hz - 45Hz
|
|
|
|
500.00mV, 5.0000V, 50.000V
|
1.5%rdg+40dgts |
Unspec'd |
|
500.00V, 1000.0V
|
Unspec'd |
|
45 Hz - 300 Hz
|
|
500.00mV
|
0.3%rdg+20dgts |
0.8%rdg+60dgts |
|
5.0000V, 50.000V
|
0.8%rdg+20dgts |
|
500.00V, 1000.0V
|
0.4%rdg+40dgts |
|
|
300Hz - 5kHz
|
300Hz - 1kHz
|
|
500.00mV,
|
0.3%rdg+10dgts |
0.8%rdg+40dgts |
|
5.0000V, 50.000V 500.00V
|
0.4%rdg+40dgts |
2.0%rdg+60dgts |
|
1000.0V
|
0.8%rdg+40dgts (300Hz--1kHz) |
1.0%rdg+40dgts |
| |
5kHz - 20kHz
|
1kHz - 20kHz
|
| 500.00mV, |
0.5%rdg+20dgts |
1dB* * |
| 5.0000V, 50.000V |
0.8%rdg+20dgts |
2dB* * |
| 500.00V |
0.5%rdg+20dgts |
3dB* * |
| 1000.0V |
Unspec'd |
Unspec'd |
| |
20kHz -- 100kHz
|
|
| 500.00mV, |
2.0%rdg+40dgts |
Unspec'd |
| 5.0000V, 50.000V |
4.0%rdg+40dgts* * |
| 500.00V |
Unspec'd |
| 1000.0V |
|
|
* From 5% to 10% of range : accuracy % of reading (or in dB) + 80d
** From 5% to 10% of range : accuracy % of reading (or in dB) +
180d
** From 10% to 15% of range : accuracy % of reading
(or in dB) + 100d
CMRR : > 90dB @ DC to 60Hz, Rs = 1k
Input Impedance : 10M
, 30pF nominal (80pF nominal for 500mV range)
Residual reading less than 50 digits with test leads shorted.
|
|
|
|