Products
|
|
DIGITAL MULTIMETER Mobile-Logger
Model KM511, KM515
|
|
 |
GENERAL SPECIFICATIONS
43000Pts Data Logger
SURGEPROTECTION6.5kV(1.2/50ms)
SPECIAL FEATURES
DISPLAY : 3-5/6 digits 5000 counts LCD display
UPDATE RATE : DIGITAL DATA : 5 per second nominal; 52 Segments Bar
graph : 60 per second nominal
LOW BATTERY : Below approx. 7V
POWER SUPPLY : Single alkaline 9V battery;
SENSING : True RMS for KM511 & KM515
|
|
APO TIMING : Idle for 17
minutes |
|
SAFETY : Meets
EN61010-1, UL3111-1, CSA C22.2 No.1010-1, & IEC1010-1 CAT III for
1000V & CAT IV 600V |
|
TRANSIENT PROTECTIONS : 6.5kV
(1.2/50 m s surge) |
|
OVERLOAD PROTECTIONS :
m A & mA : 1A/240V Fuse, IR 10kA; A : 13A/240V Fuse,
IR 10kA; V : 1050Vrms, 1450Vpeak; mV : 600VDC/AC rms; W & Other :
600VDC/AC rms |
|
POWER CONSUMPTION : 2.7
mA typical |
|
OPERATING TEMPERATURE :
0°C to 35°C, 0-80% R.H.; 35°C to 50°C, 0-70% R.H. |
|
STORAGE TEMPERATURE :
-20°C to 60°C, 80% R.H. (With battery removed) |
|
DIMENSION : (L) 186mm X
(W) 87mm X (H) 35.5mm; (L) 198mm X (W) 97mm X (H) 55mm with holster
|
|
WEIGHT : 340 gm; 430 gm
with holster Approx. |
|
SPECIAL FEATURES : Data
Hold; Backlighted display; On screen data logging 5400 points for
KM511 & 43000 points for KM515 with selectable sampling interval 0.05,
1,20, 40, 60, 120, 240, & 480 seconds |
|
ACCESSORIES : Test leads
(pair), holster, battery installed, user's manual, banana plug K type
thermocouple, Tilt stand. |
|
OPTIONAL ACCESSORIES :
PC interface Kit (RS232 optical adapter cable + software floppies)
|
|
|
ELECTRICAL FEATURES
ACCURACY IS ± (%READING DIGITS + NUMBER OF DIGITS) OR OTHERWISE
SPECIFIED, AT 23°C ± 5°C & LESS THAN 75% R.H. True RMS ACV & ACV
accuracies are specified from 5 % to 100 % of range or otherwise
specified, Crest Factor < 3:1 at full scale & < 6:1 at half scale, and
with frequency component within the specified frequency bandwidth
| Range |
KM859CF |
KM 857 |
| |
Accuracy |
| 500.00mV,5.0000V,
50.000V |
0.02% rdg + 2 dgts |
0.04% rdg + 2 dgts |
| 500.00V |
0.03% rdg + 2 dgts |
0.05% rdg + 2 dgts |
| 1000.0V |
0.1% rdg + 2 dgts |
0.05% rdg + 2 dgts |
NMRR : >
60dB @ 50/60Hz
CMRR : > 120dB @ DC, 50/60Hz, Rs = 1kW
Input Impedance : 10MW, 30pF nominal (80pF nominal for 500mV ranges)
| MEASUREMENTS :- |
| Unit |
Range |
Unit |
Range |
| Power Factor |
0.00~1.00 |
Voltage |
0~510V |
| Current |
0~4000A |
KVAR |
0~999KVAR |
| Frequency Meter |
40~70Hz |
Temperature |
0~700C |
|
|
DC VOLTAGE
|
| Range |
Accuracy* |
| 50.00mV |
0.12% rdg + 2 dgts |
| 500.0mV |
0.06% rdg + 2 dgts |
| 5.000V, 50.00V |
0.08% rdg + 2 dgts |
| 500.0V, 1000V |
NMRR : > 60dB @
50/60Hz
CMRR :
> 120dB @ DC, 50/60Hz, Rs = 1k

Input Impedance :
10M
,
16pF nominal (44pF nominal for 50mV & 500mV
ranges)
Open Circuit Voltage :
< 1.3 VDC (< 3
VDC for 50
&
500
ranges) |
|
|
| Range |
Accuracy*
|
| 50Hz - 60Hz
|
50.00mV, 500.0mV,
5.000V, 50.00V,
500.0V, 1000V |
0.5% rdg + 3 dgts |
| 40Hz - 500Hz
|
| 50.00mV, 500.0mV |
0.8% rdg + 3 dgts |
| 5.000V, 50.00V, |
1.0% rdg + 4 dgts |
| 500.0V |
| 1000V |
1.2% rdg + 4 dgts |
| Up to 20kHz |
|
| 50.00mV, 500.0mV |
0.5dB** |
| 5.000V, 50.00V, |
3dB** |
| 500.0V |
| 1000V |
Unspec'd |
CMRR :
> 60dB @ DC to 60Hz, Rs = 1K

Input Impedance :
10 M
,
16pF nominal (44pFnominal for 50mV & 500mV
ranges) |
|
DC CURRENT
|
| Range |
Accuracy*
|
Burden Voltage
|
| 500.0 m A |
0.2% rdg + 4 dgts |
0.15mV/ m A |
| 5000 m A |
0.15mV/ m A |
| 50.00 mA |
3.3mV/mA |
| 500.0 mA |
3.3mV/mA |
| 5.000 A |
0.03V/A |
| 10.00 A* |
0.03V/A |
*10A continuous, 20A for 30 second max with 5 minutes cool down interval
|
DIODE TESTER
|
| Range |
Accuracy* |
Test Current (Typical)
|
Open Circuit
Voltage |
| 2.000V |
1% rdg + 1dgt |
0.8mA |
< 3.5 VDC |
Audible Continuty Tester
Audible threshold :
Between 20
and 120
.Response time < 100 m s
|
|
AC CURRENT
|
| Range |
Accuracy*
|
Burden Voltage
|
| 50Hz – 60Hz
|
| 500.0 m A |
0.6% rdg + 3 dgts |
0.15mV/ m A |
| 5000 m A |
0.15mV/ m A |
| 50.00mA |
3.3mV/mA |
| 500.0mA |
1.0% rdg + 3 dgts |
3.3mV/mA |
| 5.000A |
0.6% rdg +3 dgts |
0.03V/A |
| 10.00A* |
0.03V/A |
| 40Hz – 1kHz |
| 500.0 m A |
0.8% rdg + 4 dgts |
0.15mV/ m A |
| 5000 m A |
0.8% rdg + 4 dgts |
0.15mV/ m A |
| 50.00mA |
0.8% rdg + 4 dgts |
3.3mV/mA |
| 500.0mA |
1.0% rdg + 4 dgts |
3.3mV/mA |
| 5.000A |
0.8% rdg + 4 dgts |
0.03V/A |
| 10.00A* |
0.8% rdg + 4 dgts |
0.03V/A |
*10A continuous, 20A for 30 second max with 5 minutes cool down interval |
|
CAPACITANCE
|
| Range |
Accuracy*
|
| 50.00nF |
0.8% rdg + 3 dgts |
| 500.0nF |
0.8% rdg + 3 dgts |
| 5.000 m F |
1.0% rdg + 3 dgts |
| 50.00 m F |
2.0% rdg + 3 dgts |
| 500.0 m F |
3.5% rdg + 5 dgts |
| 9999 m F |
5.0% rdg + 5 dgts |
*Accuracies with film capacitor or better |
|
|
| Function
|
Sensitivity (Sine RMS)
|
Range
|
| mV |
300mV |
5Hz – 125kHz |
| 5V |
2V |
5Hz – 125kHz |
| 50V |
20V |
5Hz – 20kHz |
| 500V |
80V |
5Hz – 1kHz |
| 1000V |
300V |
5Hz – 1kHz |
,Cx,diode |
300mV |
5Hz – 125kHz |
| m A,mA, A |
10% F.S. |
5Hz – 125kHz |
Accuracy : 0.01%
rdg + 2 dgts. |
|
TEMPERATURE
|
| Range
|
Accuracy
|
| -50 °C TO 1000 °C |
0.3% rdg + 3 dgts |
|
|
|
|
|
|