==================================================================  Products
 
DIGITAL MULTIMETER Mobile-Logger Model KM511, KM515

GENERAL SPECIFICATIONS

43000Pts Data Logger
SURGEPROTECTION6.5kV(1.2/50ms)

SPECIAL FEATURES
DISPLAY :
3-5/6 digits 5000 counts LCD display
UPDATE RATE :
DIGITAL DATA : 5 per second nominal; 52 Segments Bar graph : 60 per second nominal
LOW BATTERY :
Below approx. 7V
POWER SUPPLY :
Single alkaline 9V battery;
SENSING :
True RMS for KM511 & KM515


 
APO TIMING : Idle for 17 minutes
SAFETY : Meets EN61010-1, UL3111-1, CSA C22.2 No.1010-1, & IEC1010-1 CAT III for 1000V & CAT IV 600V

TRANSIENT PROTECTIONS : 6.5kV (1.2/50 m s surge)

OVERLOAD PROTECTIONS : m A & mA : 1A/240V Fuse, IR 10kA; A : 13A/240V Fuse, IR 10kA; V : 1050Vrms, 1450Vpeak; mV : 600VDC/AC rms; W & Other : 600VDC/AC rms
POWER CONSUMPTION : 2.7 mA typical
OPERATING TEMPERATURE : 0°C to 35°C, 0-80% R.H.; 35°C to 50°C, 0-70% R.H.
STORAGE TEMPERATURE : -20°C to 60°C, 80% R.H. (With battery removed)
DIMENSION : (L) 186mm X (W) 87mm X (H) 35.5mm; (L) 198mm X (W) 97mm X (H) 55mm with holster
WEIGHT : 340 gm; 430 gm with holster Approx.
SPECIAL FEATURES : Data Hold; Backlighted display; On screen data logging 5400 points for KM511 & 43000 points for KM515 with selectable sampling interval 0.05, 1,20, 40, 60, 120, 240, & 480 seconds
ACCESSORIES : Test leads (pair), holster, battery installed, user's manual, banana plug K type thermocouple, Tilt stand.
OPTIONAL ACCESSORIES : PC interface Kit (RS232 optical adapter cable + software floppies)
ELECTRICAL FEATURES

ACCURACY IS ± (%READING DIGITS + NUMBER OF DIGITS) OR OTHERWISE SPECIFIED, AT 23°C ± 5°C & LESS THAN 75% R.H. True RMS ACV & ACV accuracies are specified from 5 % to 100 % of range or otherwise specified, Crest Factor < 3:1 at full scale & < 6:1 at half scale, and with frequency component within the specified frequency bandwidth

Range KM859CF KM 857
  Accuracy
500.00mV,5.0000V, 50.000V 0.02% rdg + 2 dgts 0.04% rdg + 2 dgts
500.00V 0.03% rdg + 2 dgts 0.05% rdg + 2 dgts
1000.0V 0.1% rdg + 2 dgts 0.05% rdg + 2 dgts
NMRR : > 60dB @ 50/60Hz
CMRR : > 120dB @ DC, 50/60Hz, Rs = 1kW
Input Impedance : 10MW, 30pF nominal (80pF nominal for 500mV ranges)
 

MEASUREMENTS :-
Unit Range Unit Range
Power Factor 0.00~1.00 Voltage 0~510V
Current 0~4000A KVAR 0~999KVAR
Frequency Meter 40~70Hz Temperature 0~700C
 DC VOLTAGE
 Range  Accuracy*
 50.00mV  0.12% rdg + 2 dgts
 500.0mV  0.06% rdg + 2 dgts
 5.000V, 50.00V  0.08% rdg + 2 dgts
 500.0V, 1000V

NMRR :
> 60dB @ 50/60Hz
CMRR :
> 120dB @ DC, 50/60Hz, Rs = 1k
Input Impedance :
10M , 16pF nominal (44pF nominal for                            50mV & 500mV ranges)
 
  RESISTANCE
 Range  Accuracy*
50.00 0.2% rdg + 6 dgts
500.0 0.1% rdg + 3 dgts
5.000k , 50.00k 0.1% rdg + 2 dgts
500.0k
5.000M 0.4% rdg + 3 dgts
50.00M 1.5% rdg + 5 dgts

Open Circuit Voltage : < 1.3 VDC (< 3 VDC for 50 &
                                500 ranges)

  AC VOLTAGE

 Range  Accuracy*
 50Hz - 60Hz
 50.00mV, 500.0mV,
 5.000V, 50.00V,
 500.0V, 1000V
 0.5% rdg + 3 dgts
 40Hz - 500Hz
 50.00mV, 500.0mV  0.8% rdg + 3 dgts
 5.000V, 50.00V,  1.0% rdg + 4 dgts
 500.0V
 1000V  1.2% rdg + 4 dgts
 Up to 20kHz  
 50.00mV, 500.0mV  0.5dB**  
 5.000V, 50.00V,  3dB**
 500.0V
 1000V Unspec'd

CMRR : > 60dB @ DC to 60Hz, Rs = 1K
Input Impedance :
10 M , 16pF nominal (44pFnominal for                             50mV & 500mV ranges)

 DC CURRENT
 Range   Accuracy*  Burden Voltage
 500.0 m A  0.2% rdg + 4 dgts  0.15mV/ m A
 5000 m A  0.15mV/ m A
 50.00 mA  3.3mV/mA
 500.0 mA  3.3mV/mA
 5.000 A  0.03V/A
 10.00 A*  0.03V/A


*10A continuous, 20A for 30 second max with 5 minutes cool down interval

  DIODE TESTER
 Range  Accuracy*  Test Current  (Typical)  Open Circuit  Voltage
 2.000V  1% rdg + 1dgt  0.8mA  < 3.5 VDC

Audible Continuty Tester
Audible threshold :
Between 20 and 120 .Response time < 100 m s

 
  AC CURRENT
 Range   Accuracy*  Burden Voltage
 50Hz – 60Hz
 500.0 m A  0.6% rdg + 3 dgts  0.15mV/ m A
 5000 m A  0.15mV/ m A  
 50.00mA  3.3mV/mA  
 500.0mA  1.0% rdg + 3 dgts  3.3mV/mA
 5.000A  0.6% rdg +3 dgts  0.03V/A
 10.00A*  0.03V/A
 40Hz – 1kHz
 500.0 m A  0.8% rdg + 4 dgts  0.15mV/ m A
 5000 m A  0.8% rdg + 4 dgts  0.15mV/ m A
 50.00mA  0.8% rdg + 4 dgts  3.3mV/mA
 500.0mA  1.0% rdg + 4 dgts  3.3mV/mA
 5.000A  0.8% rdg + 4 dgts  0.03V/A
 10.00A*  0.8% rdg + 4 dgts  0.03V/A

*10A continuous, 20A for 30 second max with 5 minutes cool down interval
 CAPACITANCE
 Range  Accuracy*
 50.00nF  0.8% rdg + 3 dgts
 500.0nF  0.8% rdg + 3 dgts
 5.000 m F  1.0% rdg + 3 dgts
 50.00 m F  2.0% rdg + 3 dgts
 500.0 m F  3.5% rdg + 5 dgts
 9999 m F  5.0% rdg + 5 dgts

*Accuracies with film capacitor or better

  FREQUENCY

 Function  Sensitivity (Sine RMS)  Range
 mV  300mV  5Hz – 125kHz
 5V  2V  5Hz – 125kHz
 50V  20V  5Hz – 20kHz
 500V  80V  5Hz – 1kHz
 1000V  300V  5Hz – 1kHz
  ,Cx,diode  300mV  5Hz – 125kHz
 m A,mA, A  10% F.S.  5Hz – 125kHz

Accuracy :
0.01% rdg + 2 dgts.
 TEMPERATURE
 Range  Accuracy
 -50 °C TO 1000 °C 0.3% rdg + 3 dgts




 

        

 

  Designed & Hosted by : MID   Promoted by : GID